20 ma transmitter output – Precision Digital PD6830 User Manual

Page 9

Advertising
background image

PD6830 Pulse Input Rate/Totalizer Instruction

Manual

9

4-20 mA Transmitter Output

OUTPUT
SOURCE

Rate or total

SCALING RANGE 4.000 to 20.000 mA for any display range.

CALIBRATION

Factory Calibrated: 0.0 to 1000.0 = 4-20 mA output

UNDERRANGE

Output Underrange: 3.8 mA

OVERRANGE

Display Overrange: 22.0 mA

Output Overrange: 20.5 mA

ACCURACY

± 0.1% span ± 0.004 mA

TEMPERATURE
DRIFT

0.4 µA/

C max from 0 to 65C ambient,

0.8 µA/

C max from -40 to 0C ambient

Note: Analog output drift is separate from input drift.

EXTERNAL LOOP
POWER SUPPLY

30 VDC maximum

OUTPUT LOOP
RESISTANCE

Power supply

Minimum

Maximum

24 VDC

10

 700

30 VDC

100

 1200

Note: loop-powered backlight subtracts 150

from

maximum resistance figures above.

Advertising