Dmiiis, Frequency measurement – Amprobe DM-111S Current-Data-Logger User Manual

Page 50

Advertising
background image

DMIIIS

EN - 48

13.1.6. Measurement of harmonics

Voltage

Range

Accuracy

Resolution

DC

– 25h

26h

– 33h

34h

– 49h

(5.0%+2digit)

(10.0%+2digit)

(15.0%+2digit)

0.1V

The voltage harmonics will be null under the following threshold:
-

DC: if <1V or <2% of 1

st

harmonic

-

1

st

harmonic: if <2V

-

2

nd

49

th

: if <1V or <2% 1

st

harmonic

Current

Range

Accuracy

Resolution

DC

– 25h

26h

– 33h

34h

– 49h

(5.0%+2digit)

(10.0%+2digit)

(15.0%+2digit)

0.1A

The current harmonics will be null under the following threshold:
-

DC: if <2% of 1

st

harmonic or < 0,2% of clamp full scale

-

1

st

harmonic: if < 0,2% of clamp full scale

-

2

nd

49

th

: if<2% 1

st

harmonic or < 0,2% of clamp full scale

Setting the FLEX option the DC component will be ignored.


13.1.7. Frequency measurement

Instrument set to 50Hz

Range

Resolution

Accuracy

47

53

0,1Hz

(1.0%+1digit)

Instrument set to 60Hz

Range

Resolution

Accuracy

57

63,6

0,1Hz

(1.0%+1digit)


13.1.8. Compliance
DM-IIIS complies with the standards prescribed for:

class 2 EN61036

– static counters of active energy

class 3 IEC1268

– static counters of reactive energy


13.1.9. Temperature drift
Temperature drift: 0,1 x accuracy/K

13.1.10.

Safety

The instrument complies to the standards:

EN 61010-1 + A2(1996)

Insulation:

Class 2

Pollution:

2

Over-voltage category:

CAT III 370V~ (Phase-Earth)

CAT III 635V~ (Phase-Phase)

Advertising