Appendix 1, Messages displayed, Appendix 1 – messages displayed – Amprobe DM-111S Current-Data-Logger User Manual

Page 52

Advertising
background image

DMIIIS

EN - 50

14. APPENDIX 1

– MESSAGES DISPLAYED

Message

Description

Advices

AUTONOM:

Available memory autonomy for the recording
which is being effected

CLEAR ALL? (Enter)

The operator is trying to cancel all the recordings
effected

Press ESC in order not to cancel
the

whole

memory,

press

ENTER to confirm

CLEAR LAST?
(Enter)

The operator is trying to cancel the last recording
effected

Press ESC in order not to cancel
the last recording, press ENTER
to confirm

Data saved

The data have been saved

DATA SIZE:

Dimensions of the stored data

HOLD

By pressing the proper key, the HOLD function
has been activated

Press HOLD again to disable this
function

Password:

A recording has been started and at least 5
minutes have passed from the last activity of the
instrument (see paragraph 7).

Insert the password:
F1, F4, F3, F2

Invalid date

The inserted date is not correct

Check the inserted date

Energy Measuring

The instrument is taking an energy measurement

Press F1 to stop it

Memory Full

The memory of the instrument is full

Cancel some recordings after
transferring them to a PC

No ext supply!

A recording has been started without connecting
the external power supply

Verify it you really want to start
the recording without the external
power supply. In that case, press
START again.

No parameter sel

A recording has been started without selecting any
value to be recorded

Press START/STOP and select
at least a value entering the
MENU

No Phase selected

Voltage and/or current harmonics have been
selected and the corresponding flag has been
enabled (HARMONICS ON) but no phase voltage
or current has been selected

Select at least one phase voltage
and/or current

PASSWORD
ERROR

The inserted password is wrong (see paragraph 7)

Check the password

PASSWORD OK

The inserted password is correct

Please wait

The instrument is waiting for the recording to be
started (see paragraph 6)

Recording

The instrument is recording (see paragraph 6)

Too many param

More than 63 parameters have been selected
(harmonics included) or More than 38 parameters
with CO-GENERATION Flag enabled

Deselect some values

Too many record

The quantity of recorded data + Smp exceeds the
maximum allowed (35)

Cancel some recordings after
transferring them to a PC

ERR: SEQ

The Phase Sequence is not correct.

Check the Phase Sequence
connection.

ERR: P-

The active powers shown on the right side of the
message are negative

If there isn’t a situation of co-
generation check if the clamps
are properly connected

ERR: SEQ & P-

The active powers shown on the right side of the
message are negative and the Phase Sequence
is not correct.

If there is not a situation of co-
generation, check if the clamps
are properly connected / check
the Phase Sequence connection.

ERR: CONNECTION

The instrument has detected a wrong connection
to Voltage inputs

Check the Voltage connections
(see paragraph 11)

Error Vref

The user set a Voltage reference not compatible
with voltage at instrument’s input.

Check Voltage Reference set in
“CONFIG RECORDER”

Error1

Error 5

Error during Memory Reading

Contact Amprobe assistance