Appendix 1, Messages displayed, Appendix 1 – messages displayed – Amprobe DM-111S Current-Data-Logger User Manual
Page 52
DMIIIS
EN - 50
14. APPENDIX 1
– MESSAGES DISPLAYED
Message
Description
Advices
AUTONOM:
Available memory autonomy for the recording
which is being effected
CLEAR ALL? (Enter)
The operator is trying to cancel all the recordings
effected
Press ESC in order not to cancel
the
whole
memory,
press
ENTER to confirm
CLEAR LAST?
(Enter)
The operator is trying to cancel the last recording
effected
Press ESC in order not to cancel
the last recording, press ENTER
to confirm
Data saved
The data have been saved
DATA SIZE:
Dimensions of the stored data
HOLD
By pressing the proper key, the HOLD function
has been activated
Press HOLD again to disable this
function
Password:
A recording has been started and at least 5
minutes have passed from the last activity of the
instrument (see paragraph 7).
Insert the password:
F1, F4, F3, F2
Invalid date
The inserted date is not correct
Check the inserted date
Energy Measuring
The instrument is taking an energy measurement
Press F1 to stop it
Memory Full
The memory of the instrument is full
Cancel some recordings after
transferring them to a PC
No ext supply!
A recording has been started without connecting
the external power supply
Verify it you really want to start
the recording without the external
power supply. In that case, press
START again.
No parameter sel
A recording has been started without selecting any
value to be recorded
Press START/STOP and select
at least a value entering the
MENU
No Phase selected
Voltage and/or current harmonics have been
selected and the corresponding flag has been
enabled (HARMONICS ON) but no phase voltage
or current has been selected
Select at least one phase voltage
and/or current
PASSWORD
ERROR
The inserted password is wrong (see paragraph 7)
Check the password
PASSWORD OK
The inserted password is correct
Please wait
The instrument is waiting for the recording to be
started (see paragraph 6)
Recording
The instrument is recording (see paragraph 6)
Too many param
More than 63 parameters have been selected
(harmonics included) or More than 38 parameters
with CO-GENERATION Flag enabled
Deselect some values
Too many record
The quantity of recorded data + Smp exceeds the
maximum allowed (35)
Cancel some recordings after
transferring them to a PC
ERR: SEQ
The Phase Sequence is not correct.
Check the Phase Sequence
connection.
ERR: P-
The active powers shown on the right side of the
message are negative
If there isn’t a situation of co-
generation check if the clamps
are properly connected
ERR: SEQ & P-
The active powers shown on the right side of the
message are negative and the Phase Sequence
is not correct.
If there is not a situation of co-
generation, check if the clamps
are properly connected / check
the Phase Sequence connection.
ERR: CONNECTION
The instrument has detected a wrong connection
to Voltage inputs
Check the Voltage connections
(see paragraph 11)
Error Vref
The user set a Voltage reference not compatible
with voltage at instrument’s input.
Check Voltage Reference set in
“CONFIG RECORDER”
Error1
Error 5
Error during Memory Reading
Contact Amprobe assistance