Amprobe ACD-330T Clamp-On-Multimeter User Manual

Page 14

Advertising
background image

5. •))) / Hz

m

: Continuity, Frequency

• In the

position, press this button momentarily to toggle"-)))" continuity ON/OFF.

The continuity buzzer sounds when test value is below 100 counts (10.0

on auto

range). Pushing this button for more than 1 second will exit the continuity function
and return to the auto-ranging ohm measurement.

• Press to restart I ms PEAK hold test after entering PEAK mode.
• In the voltage and Current tests, push this button momentarily to enter dual display

mode. The small digits will indicate frequency reading. The frequency test always auto
range measuring.

6. AUTO / RANGE

m

:

• In auto-range press this button momentarily to select manual range and turn off the

"AUTO" annunciator.

• In manual range, press this button momentarily to step up 1 range at one time, press

this button for more than 1 second to enter auto-range.

• In auto-range, the " AUTO " annunciator is lit and the meter will select an appropriate

range for measurement being made. If a reading is greater than maximum available
range, " OL"(overload) is displayed on the screen. The meter selects a lower range
when reading is less than about 9% of full scale.

10