Amprobe ACD-330T Clamp-On-Multimeter User Manual

Page 35

Advertising
background image

DC VOLTAGE

ELECTRICALSPECIFICATIONS

Range Resolution Accuracy Overload Protection

400V

0.1V

1000V

1V

Accuracy is given as ±% of reading + number of least significant digits at 23º C ± 5º C,
with relative humidity Less than 80% R.H.

• Input Impedence 10M

• Input Impedence 10M

// less than 100pF

Crest factor: <3:1

• Input Impedence 10M

Specified accuracy +/- 40 digits for changes > 1 ms in duration.

1000V
AC RMS

±(1%rdg+3dgt)

AC VOLTAGE (TRUE RMS: From 10% to 100% of range.)

Range Resolution Accuracy Overload Protection

400V

0.1V

1000V

1V

1000V
AC RMS

±(1.5%rdg+3dgt)

WARNING:
The measuring duty cycle should not exceed the following limits.

0

600 A RMS

Continuous

600

700 A RMS

10 minutes ON, 10 minutes OFF

700

1000 A RMS

5 minutes ON, 20 minutes OFF

AC VOLTAGE (1 ms PEAK HOLD.)

Range Resolution Accuracy Overload Protection

400V

0.1V

1000V

1V

1000V
AC RMS

±(1.5%rdg+43dgt)

• This unit has a crest factor up to 3 on symmetric waveforms.

AC CURRENT(TRUE RMS: From 10% to 100% of range.)

Range Resolution Accuracy Maximum Overload

400 A

0.1A

400 700 A

1A

700 1000 A

1A

1000 A
RMS

±(1.2%
+5dgt

)

45

65 Hz

65

400 Hz

400
1k Hz

±(2.0%

+5dgt

)

±(4.0%
+5dgt

)

±(1.5%
+5dgt

)

±(3.0%
+5dgt

)

±(5.0%
+5dgt

)

±(2.5%

+5dgt

)

31

45Hz

400Hz

Advertising