Amprobe ACD-330T Clamp-On-Multimeter User Manual
Page 36

CURRENT (1ms PEAK HOLD)
Specified accuracy +/-40 digits for changes > 1 ms in duration.
Range
Resolution Accuracy
400A
0.1A
±(2%rdg+43dgt)
1000A
1A
±(2%rdg+43dgt)
FREQUENCYCOUNTERSENSITIVITY
INPUTRANGE
MINIMUM SENSITIVITY
(RMSSINEWAVE)
400A
3A
1000A
30A
400V
3V
1000V
30V
(Maximum input for
specified accuracy
=10 x range or1000V)
40 Hz-2 kHz
FREQUENCY(AC coupling)
Range
Resolution Accuracy Minimum Input Frequency
100Hz 0.01Hz
1kHz 0.1Hz
• Overload protection: 1000Vrms AC;< 1000000 VxHz
±(0.2%rdg+4dgt)
10Hz
DIODECHECK
Range Resolution Accuracy Test Current Test Voltage
Diode
1mV
±(1.0%rdg+2dgt) approx. 1.65 mA
<3.3V
• Overload protection: 600VRMS
RESISTANCE
Range Resolution Accuracy
Max Test Voltage Overload Protection
400
Ω
0.1
Ω
3.3V
4k
Ω
1
Ω
±(1%rdg+3dgt)
1.25V
600Vrms
• Instant continuity mode, built-in buzzer sounds when resistance is less than 10.0
Ω
.
built-in buzzer
sounds when
reading is
below approx.
100mV
AUDIBLE CONTINUITYTEST
Range Resolution Accuracy Test Current Test Voltage
approx. 1.65 mA
<3.3V
• Overload protection: 600VRMS
1mV
Diode
32