SPECIFICATIONS

General Specifications
Display count : 10000 or 4000

Measuring rate : 3 times / per second

Overrange display : “OL” or “-OL” .

Auto Power Off : Approx. 15 minutes.

Low battery indicator : is displayed. Replace the battery when the
indicator appears in the display.

Power requirement : 9V battery.

Battery life : ALKALINE 9V 50 hours (without Backlight).

Environmental Conditions

Indoor Use

Calibration: One year calibration cycle.

Operating temperature: 0 °C ~ 10 °C

11 °C ~ 30 °C (≦80% RH)

30 °C ~ 40 °C (≦75% RH)

40 °C ~ 50 °C (≦45%RH)

Storage temperature: -10 to 50 °C for current,

-20 to 60 °C for other function,

0 to 80% RH (batteries not fitted).

Temperature coefficient: 0.2 x (Specified accuracy) / °C, < 18°C, > 28°C .

Over voltage category: IEC 61010-1 600V CAT IV., 1000V CAT III.

CAT

Application field

I

The circuits not connected to mains.

II

The circuits directly connected to Low-voltage installation.

III

The building installation.

IV

The source of the Low-voltage installation.

Operating altitude: up to 2000m (6562 ft)

Conductor Size:
37mm diameter for ACD-50NAV, ACD-51NAV, ACDC-52NAV
42mm diameter for ACD-53NAV, ACDC-54NAV

Pollution degree: 2

EMC : EN 61326-1

Shock vibration: Sinusoidal vibration per MIL-T- 28800E (5 ~ 55 Hz, 3g
maximum).

Drop Protection: 4 feet drop to hardwood on concrete floor.