Amprobe ACD-50NAV ACD-51NAV ACDC-52NAV ACD-53NAV ACD-54NAV Navigator-Clamps User Manual

Page 29

Advertising
background image

23

ACDC-52NAV, ACDC-54NAV

Function

Range

Accuracy

DC A

99.99A

± (1.5% + 0.2A)

599.9A / 999.9A*

± (1.5% + 5dgt) **

AC A

0.10A ~ 99.99A

50 ~ 60Hz ± (1.5% + 5dgt) **

61 ~ 400Hz ± (2% + 5dgt) **

599.9A / 999.9A*

LPF

AC A

0.10A ~ 99.99A

50 ~ 60Hz ± (1.5% + 5dgt) **

61 ~ 400Hz ± (5% + 5dgt) **

599.9A / 999.9A*

* ACDC-52NAV: 599.9A ; ACDC-54NAV: 999.9A
** The measured value <1000dgt, add 5 dgt to the accuracy.

Overload protection: 1000Arms For ACD-53NAV, ACDC-54NAV

600Arms For ACD-50NAV, ACD-51NAV, ACDC-52NAV

Position Error: ±1% of reading.

AC Conversion Type and additional accuracy is same as AC Voltage.

AC+DC Arms Accuracy: Same as ACA spec + DCA spec.

• For better measurement accuracy of high current, do not

measurement more than 10 minutes (for ACD-50NAV, ACD-51NAV,

ACD-53NAV)

• DC A affected by the temperature and the residual magnetism.

Press HOLD key > 2 seconds to compensate it.

(3) Peak Hold : Peak MAX / Peak MIN

ACD-50NAV, ACD-51NAV, ACDC-52NAVV

Function

Range

Accuracy

AC V

140.0V

± (3.0% + 15dgt)

1400V

AC A

140.0A

± (3.0% + 15dgt)

850A

ACD-53NAV, ACDC-54NAV

Function

Range

Accuracy

AC V

140.0V

± (3.0% + 15dgt)

1400V

AC A

140.0A

± (3.0% + 15dgt)

1400A

Overload protection: 1000 Vrms
600 Arms For ACD-50NAV, ACD-51NAV, ACDC-52NVA
1000 Arms For ACD-53NAV, ACDC-54NAC

Accuracy defined for:
Sine wave, ACV>5Vrms / ACA

5Arms, Freq. 50~400Hz.

• Only suitable for the repetitive events.