Amprobe ACD-50NAV ACD-51NAV ACDC-52NAV ACD-53NAV ACD-54NAV Navigator-Clamps User Manual

Page 7

Advertising
background image

1

SYMBOLS

X

Caution ! Risk of electric
shock

Caution ! Refer to the
explanation in this Manual

,

Application around and
removal from hazardous live
conductors is permitted.

T

Equipment protected
by double or reinforced
insulation

B

Alternating Current (AC).

F

Direct Current (DC).

Temperature Measurement

e

Resistance Measurement

E

Capacitance Measurement

G

Diode Measurement

R

Continuity Beeper

J

Earth (Ground)

Y

To alert you to the presence
of a potentially hazardous
voltage, when the Meter
detects a voltage ≧30 V or a
voltage overload (OL) in V.
This symbol is displayed.

Complies with European
Directives

Conform to relevant
Australian standards

Battery

=

Do not dispose of this product as unsorted municipal waste. Contact
a qualified recycler

Safety Information

• This manual contains information and warnings that must be

followed for operating the instrument safely and maintaining the
instrument in a safe operating condition. If the instrument is used in
a manner not specified by the manufacturer, the protection provided
by the instrument may be impaired.

• The meter is intended only for indoor use.

• The meter protection rating, against the users, is double insulation

per IEC 61010-1 2nd Ed., EN 61010-1 2nd Ed., Category III 1000 Volts
AC & DC and Category IV 600 Volts AC & DC.

• Disconnect the test leads from the test points before changing the

position of the function rotary switch.

• Never connect a source of voltage with the function rotary switch in

e

,

E

, , μA, position.

• Do not expose Meter to extremes in temperature or high humidity.

• Never set the meter in

e

,

E

, , μA, function to measure the voltage

of a power supply circuit in equipment that could result in damage
the meter and the equipment under test.