BAT-500

Battery Impedance Tester

Keys And Input / Output Terminals

1

+

POWER key : Power on / off

2

READ key :

1) Press key to show the manual logged readings.
2) Press READ key again to stop reading.

3

MEMORY key :

1) Under the manual logging mode, the tester stores each single set of

logged reading to the memory by pressing MEMORY key.

2) Press and hold MEMORY key for 2 seconds to enter continuous

logging mode. Press again to stop logging.

4

V-RANGE key : Select the voltage range. (4V, 40V)

5

HOLD key :

1) Press HOLD key to freeze or unfreeze the displayed reading.

2) Press and hold HOLD key for 2 seconds to enter the interval time setting

mode for continuous data logging.

6

Ω- RANGE key : Select the impedance range. (40mΩ, 400mΩ, 4Ω, 40Ω)

7

u REL key :

1) Press u key to move the cursor to the right.
2) Press REL (Relative) key to zero the reading.

8

p key : Press p key to increase the displayed value.

9

SET key :

1) Press SET key to switch the comparator mode on or off.
2) Press and hold the SET key for 2 seconds to enter the comparator-

setting mode. Press again to store the setting in memory.

10

q key : Press q key to decrease the displayed value.

11

t

R

Key :

1) Press t key to move the cursor to the left.
2) Press

R

key to turn the beeper on or off.

12

RS-232 connector : PC interface connector.

13

– Input jack : For connecting with the black test lead plug.

14

+ Input jack : For connecting with the red test lead plug.