Measurement preparation selection, Man mode, Auto mode – Amprobe MEGATEST-5000 Megohmmeter User Manual

Page 17: Auto man

Advertising
background image

4.5 Instrument

configuration

and measurement performance

Following is a description of the parameters and way the values are set.
4.5.1 Measurement preparation selection
To prepare the measurement, press the increase or decrease key to run through the
memories until reaching the location you intend to introduce the configuration into.

NOTICE

Memorisation is possible only from cell P01 to cell P19.


4.5.2 MAN Mode
KEY OPERATION

Ш

Ч

Reach the memory cell you are interested in.


ModE

Press the MODE key repeatedly to display the MAN measuring mode.


SeT

The V2 parameter indicator is flashing.

Ш

Ч

Set the V2 voltage value you want (testing voltage). The top secondary
display (item 10 par. 4.2) will show the value you are setting.


SeT

Exit the programming phase. The flashing measuring mode advises you that
you can save the settings.


SavE

Saving the parameters in the location first selected (if requested).


StarT “Go?”

prompt displayed, PRESS START again and HOLD DOWN UNTIL

READING STABILISES.

AUTO

MAN


4.5.3 AUTO Mode
KEY OPERATION

Ш

Ч

Reach the memory cell you are interested in.


ModE

Press the MODE key repeatedly to display the AUTO measuring mode.


SeT

The V2 parameter indicator is flashing.

Ш

Ч

Set the V2 voltage value you want (testing voltage). The top secondary
display (item 10 par. 4.2) will show the value you are setting.


SeT

Exit the programming phase. The flashing measuring mode advises you that
you can save the settings.


SavE

Saving the parameters in the location first selected (if requested).


StarT “Go

?”

prompt displayed, PRESS START again to start measuring.

-15-