Accident prevention and safety measures, Uring use – Amprobe MEGATEST-5000 Megohmmeter User Manual

Page 4

Advertising
background image

1. Accident prevention and safety measures

Please read and understand these instructions BEFORE using the AMB-5KV-D.
This instrument can generate dangerously high voltages.
Only specialised personnel, well trained about electricity and its effects must use it.

When testing, the following is important:
• DO NOT TAKE MEASUREMENTS IN AN EXPLOSIVE ATMOSPHERE. i.e. GAS,

FUELS OR DUST.

• Do not take measurements in wet environments

• Do not touch exposed metal parts; i.e. alligator clips, ends of test leads, circuits, etc.,

while taking measurements

Prevent any accidental contact of yourself or others with the measuring voltage.

• Make sure the test voltage does not damage any component unable to withstand it.

The following symbols have been used in this manual:

o

Warning: comply with the instructions in the manual. Any improper use could
damage the instrument, the equipment being tested or cause personnel
injury.


p


Danger High Voltage.


t


Instrument with double insulation.

1.1 During

use

Following are two general guidelines for setting up the tests. These must be integrated
from time to time with a detailed analysis depending on the actual circumstances.

While measuring on equipment:

- Seal off the testing area and do not allow other people to approach it. If

necessary, put up a sign.

- Have enough room available, plus properly insulated tables or supports.
- Take note of which parts could be live should there be defective insulation.
- Keep the area and the measuring cables clean from dust, oil or other foreign

matter.

- On a routine basis, check for any abrasion or breakage on the measuring

cables.

- Never exceed the maximum testing voltage for which the tested device was built

and/or rated.

-2-

Advertising