Single-Phase Power & 3-Phase Balanced-Load Power Functions

Set the slide-switch function-selector to the “3~Bal 1~” Power position.

• Default at last selected function.

• Press “3~Bal 1~” button momentarily to toggle between “Single-

Phase” and “3-Phase Balanced Load” Power functions. Annunciators “

” and “3~” turn on respectively.

• Press SELECT button momentarily selects between W (real power), VAR

(reactive power), VA (apparent power) & kWHr (real-time readings or
stored result) functions. In W (real power), VAR (reactive power), or VA
(apparent power) function:

1. PF (Total Power Factor) is displayed automatically in the secondary mini

display.

2. Annunciator “A-lags-V” turns on to indicate an inductive circuit is

being measured. That is, the Current waveform is lagging the Voltage
waveform, and the phase-shift angle θ is “+”.

3. On the contrary, together with significant PF values, WITHOUT

A-lags-V” being turned on indicates that a capacitive circuit is being
measured. That is, the Current waveform is leading the Voltage
waveform, and the phase-shift angle θ is “-”.