Amprobe ACD-45PQ Power-Quality-Clamp User Manual
Page 26
20
Ohms
Range
Accuracy
5Hz ~ 500Hz
0.5%+4d
Open Circuit Voltage: 0.4VDC typical
Audible Continuity Tester
Audible threshold: between 10 and 300.
Response time: 250µs
Single-Phase & 3-Phase Balanced-Load Power
RANGE 5) 0 ~
600.0kVA
Accuracy 1) 2) 3)
F ~ 10th
11th ~ 45th
46th ~ 51st
@ PF = 0.99
~ 0.1
2.0%+6d
3.5%+6d
5.5%+6d
RANGE 5) 0
~ 600.0kW /
kVAR
Accuracy 1) 2) 4)
F ~ 10th
11th ~ 25th
26th ~ 45th
46th ~ 51st
@ PF = 0.98 ~
0.70
2.0%+6d
3.5%+6d
4.5%+6d
10%+6d
@ PF = 0.70 ~
0.50
3.0%+6d
@ PF = 0.50 ~
0.30
4.5%+6d
@ PF = 0.30 ~
0.20
10%+6d
15%+6d
1. Specified accuracy is for ACA clamp measurement at the center of jaws.
When the conductor is not positioned at the jaw center, position errors
introduced are:
• Add 1% to specified accuracy for ACA measurements made WITHIN
jaw marking lines (away from jaw opening)
• Accuracy is not specified for ACA measurement made BEYOND jaw
marking lines (toward jaws opening)