Amprobe ACD-45PQ Power-Quality-Clamp User Manual

Page 27

Advertising
background image

21

2. Add 4d to specified accuracy for 3-Phase Balanced-Load Power

measurements.

3. Add 1% to specified accuracy @ ACA fundamental < 6A or ACV

fundamental < 90V. Accuracy is not specified @ ACA fundamental < 1A
or ACV fundamental < 30V

4. Add 1% to specified accuracy @ ACA fundamental < 6A or ACV

fundamental < 90V. Accuracy is not specified @ ACA fundamental < 2A
or ACV fundamental < 50V

5. 0 ~ 360.0

Total Power Factor (PF)

Range

Accuracy 1)

0.10 ~ 0.99

F ~ 21st

22nd ~ 51st

3d

5d

1) Specified accuracy @ ACA fundamental > 2A ; ACV fundamental > 50V

A-lags-V Indication:

LCD annunciator “A-lags-V” turns on to indicate an inductive circuit, or
Current A lags Voltage V (i.e., phase-shift angleθ is “+”).

A-lags-V Indication is specified at 50/60Hz fundamental without the presence
of harmonics, and at ACV > 90V, ACA > 9A and PF < 0.95

kWHr (kilo-Watt-Hour Energy)

Time base accuracy: < 30ppm

Non-volatile memory: Separately stores one 3-Phase-Balanced-Load and one
Single-Phase result

3-Phase Unbalanced-Load Power

This 3-Phase Unbalanced-Load Power measurement is achieved thru the
calculation of discrete single-phase measurements that are taken one at a time
manually. Since it is not real-time on all 3 phases simultaneously, it is intended
only for stable power conditions without significant power fluctuations over
the time of measurements. Result accuracy is hence the accumulated accuracy
of the discrete single-phase measurements plus the associated fluctuations.