Amprobe ACD-45PQ Power-Quality-Clamp User Manual
Page 25
19
ACA Current (Clamp-on)
Range
Accuracy 1) 2)
50Hz / 60Hz
40.00A, 400.0A, 600A
1.0% + 5d
45Hz ~500Hz
40.00A, 400.0A
2.0% + 5d
600A
2.5% + 5d
500Hz ~ 3.1kHz
40.00A, 400.0A
2.5% + 5d
600A
3.0% + 5d
ACA AutoVATM Threshold: 1A AC (40Hz ~ 500Hz only) nominal
Crest Factor:
< 3 : 1 at full scale & < 6 : 1 at half scale for 40.00A, 400.0A & 600A ranges
1. Induced error from adjacent current-carrying conductor: < 0.06A/A
2. Specified accuracy is from 1% to 100% of range and for measurements
made at the jaw center. When the conductor is not positioned at the jaw
center, position errors introduced are:
• Add 1% to specified accuracy for measurements made WITHIN jaw
marking lines (away from jaw opening)
• Add 4% to specified accuracy for measurements made BEYOND jaw
marking lines (toward jaws opening)
PEAK-rms HOLD (ACA & ACV only)
Response: 65ms to >90%
Frequency
Range
Accuracy
5Hz ~ 500Hz
0.5%+4d
Sensitivity (Sine RMS)
40A range: > 4A
400A range: > 40A
600A: > 400A
600V range: > 30V