Channel mode register, Channel relay control register – HP E1418A User Manual

Page 137

Advertising
background image

Channel Mode

Register

The Channel Mode Register reports or programs the voltage or current
operating mode for each channel.

Address

Base + 1A

16

Bit

15

14

13

12

11

10

09

08

07

06

05

04

03

02

01

00

READ

Ch

16

Mode

Ch

15

Mode

Ch

14

Mode

Ch

13

Mode

Ch

12

Mode

Ch

11

Mode

Ch

10

Mode

Ch

09

Mode

Ch

08

Mode

Ch

07

Mode

Ch

06

Mode

Ch

05

Mode

Ch

04

Mode

Ch

03

Mode

Ch

02

Mode

Ch

01

Mode

WRITE

Ch XX Mode: Writing a one (1) to this field selects the voltage output
mode for that channel. Writing a zero (0) to this field selects the current
output mode for that channel.

The mode program jumpers on the terminal module override values written
to this register. When the P/J jumper for a channel is set to jumper
selectable, writing to this register has no effect.

Reading this register returns the channel mode. The channel mode is
reported in this register for both VXIbus backplane programming and
jumper programming on the terminal module.

Note

Channels 09 through 16 will return ones (1) if the module is configured as
an 8-channel device.

Channel Relay

Control Register

The Channel Relay Control Register programs and reports the state of each
channel relay.

Address

Base + 1C

16

Bit

15

14

13

12

11

10

09

08

07

06

05

04

03

02

01

00

READ

Ch

16

Relay

Ch

15

Relay

Ch

14

Relay

Ch

13

Relay

Ch

12

Relay

Ch

11

Relay

Ch

10

Relay

Ch

09

Relay

Ch

08

Relay

Ch

07

Relay

Ch

06

Relay

Ch

05

Relay

Ch

04

Relay

Ch

03

Relay

Ch

02

Relay

Ch

01

Relay

WRITE

Ch XX Relay: Writing a one (1) to this field opens the channel relay.
Writing a zero (0) to this field closes the channel relay. The write operation
will complete before the relays change state. The relays take approximately
5 milliseconds to change state.

Reading this register returns a one (1) in each field if the channel relay is
programmed open and a zero (0) if the channel relay is programmed closed.

Note

Channels 09 through 16 will return ones (1) if the module is configured as
an 8-channel device.

136 HP E1418A Register-Based Programming

Appendix B

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Advertising