DIAGnostic

The

DIAGnostic

subsystem contains commands used to check the module

configuration, enable or disable all outputs simultaneously, change a channel’s
mode without opening and closing the output relay, and check the state of the trigger
on all channels.

Subsystem Syntax

DIAGnostic

:CALibration

:OUTPut

[:MODE] <

path

>

[:MODE]?

:CONFigure?
:OUTPut

:ALL

[:STATe] <

state

>

:SOURce

n

:FUNCtion

:MODE <

mode

>

:TRIGgered?

DIAGnostic:CALibration:OUTPut[:MODE]

DIAGnostic:CALibration:OUTPut[:MODE] <

path

>

sets the voltage calibration

path for the module to either the CAL output terminals or at each output channel.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

<

path

>

discrete

CALBus

or

CHANnel

none

Comments

•

The typical calibration mode for the module is through the CAL output terminals.
Following

*RST

or power-on, the default calibration mode is through the CAL

output terminals (and this command is not needed).

•

This command with the

CHANnel

parameter, sets calibration to be performed at

each channel’s output terminals (and so includes the channel output relay
contacts).

•

This command must be sent before any

CALibration

n

:CONFigure:VOLTage?

or

CALibration

n

:CONFigure:CURRent?

query.

•

*RST State: Calibration paths set to use the CAL output terminals.

DIAGnostic

76 HP E1418A SCPI Command Reference

Chapter 3