Source n :current[:level][:immediate][:amplitude, Source – HP E1418A User Manual

Page 86

Advertising
background image

SOURce

n

:CURRent[:LEVel][:IMMediate][:AMPLitude]

SOURce

n

:CURRent[:LEVel][:IMMediate][:AMPLitude] <

value

| MIN | MAX | DEF>

immediately outputs the current set in value at channel n when the output relay is closed.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

SOURce

n

keyword

1 to 8 or 1 to 16

1

<

value

>

numeric (float 32) or

discrete

-0.02 Amps to +0.02 Amps or

MIN | MAX | DEF

0

Comments

•

In the keyword,

SOURce

n, the valid range of n is:

1 to 8 for

E1418A, E1418A Option 002

1 to 16 for

E1418A Option 001, E1418A Option 003.

Values of n outside these ranges will generate an error.

•

The

<

value

>

parameter is a floating point 32-bit number in the range of -0.02 A

to +0.02 A.

MIN

sets -0.02 Amps.

MAX

sets +0.02 Amps.

DEF

sets 0 Amps.

•

This command immediately programs the specified current on channel n. The
output state of the channel must be enabled for the current to appear at the output
terminals.

•

If the channel is in the voltage mode when this command is received, an error is
generated.

•

Related Commands:

SOURce

n

:CURRent[:LEVel][:IMMediate][:AMPLitude]?

,

SOURce

n

:FUNCtion:MODE CURRent

,

OUTPut

n

[:STATe]

SOURce

Chapter 3

HP E1418A SCPI Command Reference 85

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com