1st Value Returned

Indicates an 8-channel or 16-channel configuration.

Bit

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

0

Use

Not Used

Channels

Channels can have one of two values; all 1’s (decimal 7) for an 8-channel
configuration, and all 0’s (decimal 0) for a 16-channel configuration.

2nd Value Returned

Indicates the type of Terminal Module (if any) installed. See page 77 for
additional information.

3rd Value Returned

Indicates isolated or non-isolated plug-on modules for each channel.

Bit

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

0

Use Ch16 Ch15 Ch14 Ch13 Ch12 Ch11 Ch10 Ch9

Ch8

Ch7

Ch6

Ch5

Ch4

Ch3

Ch2

Ch1

When a bit is set to a 0 an isolated plug-on is present. A bit set to a 1
indicates a non-isolated plug-on. For 8-channel devices, channels 9 through
16 bits are set to 1.

4th Value Returned

Indicates the state of the V/I jumper on the terminal module as either
current or voltage when the P/J jumper is in the jumper selectable position.
See page 77 for additional information.

5th Value Returned

Indicates the state of the channel output relay. See page 77 for additional
information.

6th Value Returned

Indicates the state of the P/J jumper on the terminal module. See page 77
for additional information.

Changing

Configuration

Configuration changes are made using Hewlett-Packard add-on kits. These
procedures support the following add-on kits:

Kit Part Number

Description

Use

HP E1523A

1-Channel Isolation

Plug-On Module

Change one or more channels

from Non-Isolated to Isolated

Output.

HP E1524A

8-Channel

Non-Isolated

Expansion Kit

Add 8 Channels of Non-Isolated

Output.

HP E1525A

8-Channel Isolated

Expansion Kit

Add 8 Channels of Isolated

Output.

167 Configuration and Disassembly