Calibration

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

8-5

8.4 Performing Calibration Trim

Introduction

For most applications entering the cal factor for each cell will achieve satisfactory system
performance. However, it is possible to perform a calibration trim procedure in which the
Analyzer/Controller and cell combination are used to measure a reference solution of known
conductivity; the reading of the Analyzer/Controller is adjusted to match.

The same procedure can be used to adjust the Analyzer/Controller’s reading while the cell is in
the process, if a reference instrument is used to determine the process conductivity. In this case,
the process fluid becomes the “reference solution”.

Materials and procedure

To perform calibration trim using a standard reference solution, follow the instructions in Table
8-3.

Materials required are:

A reference solution of known conductivity near the point of interest, with the temperature
controlled (or measured and compensated) to within

±

1

°

C. Conductivities of potassium

chloride solutions are provided in Table 8-2. Solutions must be prepared with high-purity
de-ionized, CO

2

-free water, and dried potassium chloride.

A container for the reference solution, large enough to immerse the cell to measuring depth.

De-ionized water to rinse the cell.

Table 8-2 Conductivity of Potassium Chloride Solutions at 25

°

C

* M = Molarity; 1M = 74.555g potassium chloride per liter of solution

Concentration M*

Conductivity (microsiemens

per cm)

0.001

147.0

0.005

717.8

0.01

1,413

0.02

2,767

0.05

6,668