Voltage & current output terminals, Relay terminals (optional), Figure 6-1 terminal connections – Honeywell 9782 Series User Manual

Page 88

Advertising
background image

Conductivity/Resistivity Analyzer/Controller

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

6-6

9

SH

G

R

G

R

W

K

W

K

NC

C O M

N O
N C

C O M
N O

L1 (K)

L 2 / N

Relay 2
Note 1

Relay 1
Note 1

120V/240V
15A
50-60 Hz
Note 5

Power &
Relay Terminals

No. 2 Temp.

Compensator

No. 1 Temp.

Compensator

Cell 2

Electrodes

Cell 1

Electrodes

Note 4

Case (as viewed from front)

Note 3

NO

COM
NC

NO

COM

NC

8

7

6

5

4

3

2

1

Voltage & Current

Output Terminals

+

-

+10

VC

+1

mA

I(+)

I(-)
+10V 2000 min
V C o m

+ 1 V 2 0 0 m i n

4-20 mA 600 min

NO
COM
NC

NC

COM

NO

Relay 3
Note 6

Relay 4
Note 6

Relay Terminals
(optional)

Two-Cell Cat. 9782 with Output/relays

9

8

7

6

5

4

3

2

1

SH

G

R

G

R

W

K

W

K

NC

Com

NO

NC

Com
NO
L1(K)
L2/N

Relay 2
Note 1

Relay 1
Note 1

120V/240V 15A
50-60 Hz
Note 5

Power &
Relay Terminals

No. 2 Temp.

Compensator

No. 1 Temp.

Compensator

Cell 2

Electrodes

Cell 1

Electrodes

Note 4

Case (as viewed from front)

Note 3

1

2

3

4
5

6

O U T 1 -

O U T 1 +

O U T 2 -

O U T 2 +

O U T 3 -

O U T 3 +

Output Board
Connections
(4-20 mA)

Two-Cell Cat. 9782 - MultiOptions

K- black

W- white

G- green

R- red

B- blue

N- brown

O- orange

Y- yellow

A- gray

V- violet

P- pink

S- slate

T- tan

Color Code

a/n 23177

NOTES:
1. Relay 1 & 2 contacts max. ratings
(Resistive Load)
Current: 2A AC or DC
Voltage: 250 VAC, 220 VDC
Switching Power: 125VA, AC: 60 W, DC.
2. See Cell Connection Diagram or directions
for wiring restrictions.
3. Coaxial cable used with greater than 20 ft.
cell/Analyzer seperation with Standard
Range instruments only.
4. For pure water samples in plastic piping,
ground the black cell electrode lead near
the cell. Alternately, connect to the 9782
ground screw as shown dotted. DO NOT
ground 10, 25 or 50 constant cells.
5. Analyzer is factory set for 120V operation.
For 240V operation, see "Power
Connections" in this manual.
6. Relay 3 & 4 contact ratings 3A @ 120V AC/
28 VDC resistive.

Figure 6-1 Terminal Connections