Table 5-1 procedure for performing i/o setup – Honeywell 9782 Series User Manual

Page 62

Advertising
background image

Conductivity/Resistivity Analyzer/Controller

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

5-6

Table 5-1 Procedure for Performing I/O Setup

Step

Screen

Action

1

MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP

DIAGNOSTICS
| PREV | NEXT

Go to I/O SETUP.

2

RELAY SETTINGS

RLY1 RLY2 RLY3 RLY4
ACLN RNGS ALM3 ALM4

DIAGNOSTICS
| PREV | NEXT

The first screen shows the present use of available relays.
This information is read-only.

In this example the 9782 has four relays. Relay 1 is used
for AutoClean. Relay 2 is used for indication of automatic
output range switching. Relay 3 is used to trigger an alarm
annunciator for Alarm 3. Relay 4 is used for Alarm 4
annunciation.

Go to the next screen.

3

ANALOG OUTPUTS

OUTPUT1 C1
OUTPUT2 C2
OUTPUT3 RATO

DIAGNOSTICS
| PREV | NEXT

The next screen shows the present assignments for the
optional current or voltage outputs, if available. This
information is read-only.

1

Go to the next screen.

4

ATTENTION
RECONFIG OF ANY I/O
SETUP PARM WILL
RESET CONFIG PARMS
TO FACTORY DEFAULTS
MENU TO EXIT
DIAGNOSTICS
| PREV | NEXT

I/O setup is normally performed only once, before field
connections are made to the Analyzer/Controller. You can
also use this menu to view parameters. If you do not
change anything, you will not cause a cold reset of the
unit. You can exit now using the MENU key. If you go
past this “Attention” screen, you must go through the entire
sequence of I/O setup screens before returning to the
main menu.

Go to the next screen.

5

ADVANCED FEATURES

AUTOCLEAN YES
RANGE STATUS NO

DIAGNOSTICS
| | NEXT

Specify whether AutoClean should be enabled. If YES, a
relay will be dedicated to this use.

Specify whether you want to use a relay to indicate when
the range shifts for the analog output based on Cell 1’s
input. (Output scaling is specified in Step 13.) If YES, a
relay will be assigned to indicate when the Cell 1 range
switch occurs.

Go to the next screen.

1

If the system uses a DE option card, no analog outputs are supported by the hardware. The screen will show “DE

ENABLED”, instead of displaying output assignments.