Honeywell 9782 Series User Manual

Page 21

Advertising
background image

Introduction

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

1-7

Selecting a parameter for edit

To select a parameter for edit:

display the screen containing the parameter

use the UP or DOWN keys to highlight the parameter name

press the ENTER key to highlight the displayed current value

Editing a parameter assignment from a list of available choices

To edit a parameter having a text string as an assigned value:

select the parameter and highlight its current value as described above

use the UP or DOWN keys to display the other valid choices

when your choice is displayed, press ENTER

Editing a parameter numeric value

To edit a parameter having a numeric value:

select the parameter and highlight its current value as described above; the “DPP”,

and

soft keys will be displayed (During calibration the

and

soft keys will be

displayed once the UP or DOWN key has been pressed.)

use the “DPP” soft key to position the decimal point

use the

or

soft keys to move the cursor to the digit to be changed

use the UP and DOWN keys to increment or decrement the value

when all digits necessary have been changed, press ENTER

Controlling an AutoClean operation manually

To initiate an AutoClean operation manually, press the DISPLAY key until the AutoClean
screen is displayed. Next press the “START” soft key.

This display can also be used to see how much time is remaining until the next scheduled
AutoClean operation. When AutoClean is active, the alarm stripe will display “AUTO CLEAN”,
and the AutoClean display will show 00:00:00.

If necessary, pause the operation using the “PAUSE” soft key. To resume the operation, press
“PAUSE” again. To cancel the operation, use the “STOP” soft key. If pause is left on, the
sequence will resume automatically after twenty minutes.

At the start of the AutoClean operation alarms and outputs will be held at current levels,
regardless of whether the AutoClean was scheduled or started manually. The alarms and outputs
will be held throughout the AutoClean operation.

Advertising