Performance under reference operating conditions

Display Accuracy:

Standard Ranges (Conductivity/Resistivity); ±0.5% of reading.

Wide Ranges (Conductivity); ±0.5% of reading or 0.02% of span.

Concentration; ±0.5% of span, ±1% conformity to conductivity vs. concentration data tables.

Output accuracy: Adjustable to within 0.01 mA, repeatable to within ±0.1% of span.

Reference Operating Conditions: 25 ±1

°

C; 10 to 40

%

RH; 120 or 240 Vac; process temp 25

°

C

measured

Operating Influences under normal operating conditions

Effect on accuracy (

%

of span): Temperature: 0.05

%

per

°

C; RH:

<

1

%

; Line Voltage:

<

0.1%

per volt; Power Loss: memory retention by EEPROM (no batteries).

2.1.7 Outputs

Standard alarm/control relays

Two SPDT Form C general purpose relays rated 0.6 amps at 120/240 Vac, 0.6 amps at 110 Vdc,
2 amps at 30 Vdc. Maximum switching power for ac, 125 VA resistive; for dc, 60W resistive.

Optional alarm/control relays

9782C- _ _ - R2:

Two additional SPDT Form C, rated at 3 amps at 120 Vac

or 28 Vdc, no analog outputs
9782C- _ _ - R0:

Two hermetically sealed Form C, rated at 3amps at 120 Vac

or 28 Vdc, plus one analog output 0 to 1 Vdc or 0 to 10 Vdc
or isolated 4 to 20 mA.

Optional analog outputs

One, two or three 4 to 20 mA outputs are available via an optional plug-in circuit card and
program module. Isolated from ground, inputs and one another.
Maximum isolation voltage: 240 Vac (will withstand 1500V hi-pot test).
Output adjustment: to within 0.01 mA.
Repeatability: ±0.1% of span. 600 ohm maximum load resistance. Proportional to selected
output
range.

9782C- _ _ - 00:

None

9782C- _ _ -VC:

One analog output: 0 to1V or 0 to10 Vdc or isolated 4 to20

mA dc (600 ohms maximum load)
9782C- _ _ - C2:

Two analog outputs: both isolated 4 to 20 mA dc.

9782C- _ _ - C3:

Three analog outputs: all 4 to 20 mA dc

One analog output also available in combination with optional relays (see above).