Conductivity/Resistivity Analyzer/Controller

Appendix D - 2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

Example 1: To determine the check resistor value needed to simulate conductivity measurement
of 10

µ

S, use cell constant 0.1 and perform the following calculation:

10 k ohms = (0.1) x (1,000,000)
10

Example 2: To determine the check resistor value needed to simulate resistivity measurement of
10 M ohms, use cell constant 0.01 and perform the following calculation:

100 K ohms = (0.01) x (10,000,000)

Concentration values

If a wide range model is used for concentration measurement, obtain the appropriate check
resistance value from the table below.

Data for Concentration Range Measurements

Material/Weight % Concentration

Simulation Resistance (ohms) @ 25º C

Cell Constant

10

25

50

Hydrochloric Acid (HCl)
0
1
4

242.5

68.9

485.0
137.7

Sulfuric Acid (H2SO4)
0
1
4

215.5

56.0

538.7
140.0

1077.4

280.0

Sodium Chloride (NaCl)
0
1
4

574.1
195.2

1435.1

398.0

2870.3

796.1

Sodium Hydroxide (NaOH)
0
1
4

189.2

54.0

473.0
135.1

946.1
270.1