Fluke NetDAQ 2645A User Manual

Page 191

Advertising
background image

Maintenance

Performance Test

6

6-19

Resistance Range*

Decade Resistor

Minimum Reading

Maximum Reading

30 k

Short Circuit (Zero)

700

1 k

30 k

29 k

29.681 k

30.019 k

300 k

290 k

289.07 k

292.63 k

3 M

2.9 M

2.8607 M

2.9410 M

* The resistance accuracy in this table makes allowance for up to 0.1 Ohm of lead wire
resistance plus 0.01% decade resistance tolerance. You must add any additional lead wire
resistance present in your setup to the resistance values given in this table.

Resistance Range*

5700A

Minimum Reading

Maximum Reading

30 k

Short Circuit (Zero)

700

1 k

30 k

19 k

19.686 k

20.014 k

300 k

190 k

189.60 k

192.10 k

3 M

1.9 M

1.8740 M

1.9277 M

* The resistance accuracy in this table makes allowance for up to 0.1 Ohm of lead wire
resistance. You must add any additional lead wire resistance present in your setup to the
resistance values given in this table.

5.

Close Spy Window.