Error Messages & Exception Conditions

Error and Status Messages

J

J-23

No instruments are available for association!

You tried to create an instrument icon from the Available Instruments list when
there were no instruments listed.

Instruments on the network are either shown as icons on the Icon Bar or are listed
on the Available Instruments list. Adding a new instrument to the network and
Icon Bar is a 2-step process. First, you add the instrument to the Communications
Configuration File, where all the network parameters are saved (select the
Communications Config command from the Setup menu). The new instrument
appears on the Available Instruments list. Second, you create an instrument icon
(select the Create Instrument Icon command from the Setup menu) by selecting
the instrument from the Available Instruments list. When an instrument is placed
on the Icon Bar, it is removed from the Available Instruments list. When an
instrument is removed from the Icon Bar, it is returned to the Available
Instruments list.

No timers available! Close another application and try again!

Windows timer resources have been exceeded.

Close other applications to free Windows timers and continue operations.

Not responding to command!

Your instrument has detected a fault during internal performance monitoring.

This message does not appear under normal operating conditions. Note the
message and click Yes to continue operations or No to terminate operations. If the
fault recurs, stop operations, and cycle instrument power. If the fault still persists,
remove the instrument from service and return it to a Fluke Service Center for
repair, along with a description of the error message.

Only [Number] bytes free disk space remain on [Drive] for Trend Link
data!

You are logging to a Trend Link file and the minimum Low Disk Space amount
has been reached.

When you are logging to a Trend Link data file, NetDAQ Logger periodically
checks for available disk space. When this space reaches the value set in the Auto
Rollover dialog box, this warning is displayed and a rollover of the Data Set file
automatically takes place. To change the disk space default value (1000 k bytes),
refer to Low Disk Space for Trend Link section of the Auto Rollover dialog box.