2-59

Table 2-4. Network Error Messages (cont)

Reported Message

Comment

Check

2. Try Verify Communications. If it still

fails, then either the host computer
Ethernet adapter is not working or the
instrument is defective. If binding
errors occurred during booting, see
Appendix I for more information. Also
refer to "Error and Status Messages"
in Appendix J.

Connection is down!

The host computer
cannot establish
network communication.

1. Make sure the Ethernet adapter driver

and TCP/IP software are installed.
See “Installing Host Computer
Networking Software.”

2. Make sure the host computer IP

address and subnet mask are set
properly.

3. For an isolated network installation on

Windows 95 or Windows NT, the host
computer IP address must be
198.178.246.1xx.

For a general network installation, you

may need to set a default gateway
address also. See “Setting Host
Computer Networking Parameters”
earlier in this chapter.

Unable to initialize
Winsock!

The host computer
software cannot find or
open the Winsock DLL.

Make sure the TCP/IP and Winsock
software are installed. See “Installing Host
Computer Networking Software” in this
chapter.