Fluke NetDAQ 2645A User Manual

Page 329

Advertising
background image

Error Messages & Exception Conditions

Error and Status Messages

J

J-31

Unable to load Trend Link DLL!

NetDAQ Logger is unable to load the Trend Link

c:\Program

Files\Fluke\tl\tbase.dll

into memory.

Check that you have sufficient memory space to load

tbase.dll

, that the

tbase.dll

has not been corrupted, or you are not trying to operate with

portions of an older version of Trend Link. Reinstall Trend Link following the
procedures in Chapter 2.

Unable to open Communications Configuration File!

You tried to open the Communications Configuration File but failed.

Check the read/write permissions for the

c:\Program

Files\fluke\netdaq\ccf.cfg

file. This error can be caused by having

this file Read-Only. Remove the Read-Only property from the file. Other causes
include the

ccf.cfg

file does not exist or is corrupted. In either of these cases,

close the application and delete the current

ccf.cfg

file. Then restart the

application and select the Communications Config command from the Setup menu
to create a new

ccf.cfg

file.

Unable to open communications socket with instrument!

You tried to communicate with the instrument and the Socket Port could not be
opened.

The Windows resources are out of memory or there is a configuration problem
with the networking software. For example, if you accidentally close the Winsock
icon you will receive this message. Close all applications and restart Windows or
close NetDAQ Logger and try again. Also check your networking software.

Unable to Open Mx+B File!

You have specified an Mx+B file that is not readable.

Make sure the file is a text file, and that it resides in a directory with the proper
read permissions.

Unable to open RS232 port!

Your host computer RS-232 port cannot be opened.

Check your host computer RS-232 port operation. Your COM port may be
disabled for an internal modem or other device or may be in use by another
application. The RS-232 port is used only for calibration procedures.

Advertising