2 cautions 4, 2 cautions – Fluke 9107 User Manual

Page 11

Advertising
background image

DO NOT REMOVE INSERTS AT HIGH TEMPERATURES. Inserts

will be the same temperature as the display temperature. Use extreme care
when removing hot inserts.

DO NOT operate this unit without a properly grounded, properly polar-

ized power cord.

DO NOT connect this unit to a non-grounded, non-polarized outlet.
HIGH VOLTAGE is used in the operation of this equipment. SEVERE

INJURY OR DEATH may result if personnel fail to observe safety pre-
cautions. Before working inside the equipment, turn power off and dis-
connect power cord.

• Always replace the fuse with one of the same rating, voltage, and type.
• Overhead clearance is required. DO NOT place unit under a cabinet or

other structure.

DO NOT use this unit for any application other than calibration work.
DO NOT use this unit in environments other than those listed in the

user’s manual.

DO NOT turn the unit upside down with the inserts in place; the inserts

will fall out of the unit.

DO NOT operate near flammable materials.
• Use of this instrument at HIGH TEMPERATURES for extended periods

of time requires caution.

• Completely unattended high temperature operation is not recom-

mended for safety reasons.

• Before initial use, after transport and anytime the dry-well has not been

energized for more than 10 days, the calibrator must be energized for a
dry-out period of 1 to 2 hours before it can be assumed to meet all of the
safety requirements of the IEC1010-1.

• Follow all safety guidelines listed in the user’s manual.
• CALIBRATION EQUIPMENT should only be used by TRAINED PER-

SONNEL.

1.2.2

Cautions

To avoid possible damage to the instrument, follow these guidelines.

• Components and heater lifetime can be shortened by continuous high

temperature operation.

4

Hart Scientific

1 Before You Start

Advertising
This manual is related to the following products: