2 safety information 2, 1 warnings 2, Safety information – Fluke 9107 User Manual

Page 9: 1 warnings, 2 safety information

Advertising
background image

Symbol

Description

Canadian Standards Association

OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 refers to the
level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATE-
GORY II is energy-consuming equipment to be supplied from the fixed installation. Exam-
ples include household, office, and laboratory appliances.

C-TIC Australian EMC

1.2

Safety Information

Use the instrument only as specified in this manual. Otherwise, the protection
provided by the instrument may be impaired. Refer to the safety information
below and throughout the manual.

The following definitions apply to the terms “Warning” and “Caution”.

• “Warning” identifies conditions and actions that may pose hazards to the

user.

• “Caution” identifies conditions and actions that may damage the instru-

ment being used.

1.2.1

Warnings

To avoid possible electric shock or personal injury, follow these guidelines.

BURN HAZARD Do Not touch the well access surface of the unit.

(Refer to Figure 1 on page 3.)

• The temperature of the well access (1) will be the same as the actual tem-

perature shown on the display (2), e.g. if the unit is set at 600°C and the
display reads 600°C, the well is at 600°C.

• The top sheet metal (3) of the dry-well may exhibit extreme temperatures

for areas close to the well access.

• The air over the well can reach temperatures greater than 200°C.

PROBES SHOULD ONLY BE INSERTED AND REMOVED FROM
THE UNIT WHEN THE UNIT IS SET AT TEMPERATURES LESS
THAN 200°C.

DO NOT TURN OFF THE UNIT AT TEMPERATURES HIGHER

THAN 100°C. This could create a hazardous situation. Select a set-point
less than 100°C and allow the unit to cool before turning it off.

2

Hart Scientific

1 Before You Start

CAT II

Advertising
This manual is related to the following products: