• Components and heater lifetime can be shortened by continuous high

temperature operation.

• Most probes have handle temperature limits. Be sure that the probe handle

temperature limit is not exceeded in the air above the unit.

• (9107 only) Always use the well insulator, see Section 5.4.
• Allow for test probe expansion inside the well as the dry-well heats.
DO NOT use fluids to clean out the well.
• Never introduce any foreign material into the probe hole of the insert.

Fluids, etc. can leak into the calibrator causing damage.

DO NOT change the values of the calibration constants from the factory

set values. The correct setting of these parameters is important to the
safety and proper operation of the calibrator.

DO NOT slam the probe stems in to the well. This type of action can

cause a shock to the sensor and affect the calibration.

DO use a ground fault interrupt device.