10 test probe calibration 51, 1 comparison methods 51, 1 direct calibration 51 – Fluke 9107 User Manual

Page 55: 2 comparison calibration 51, 10 test probe calibration, 1 comparison methods, 1 direct calibration 10.1.2 comparison calibration

Advertising
background image

10

Test Probe Calibration

For optimum accuracy and stability, allow the calibrator to warm up for 10
minutes after power-up and then allow adequate stabilization time after reach-
ing the set-point temperature. After completing calibration, allow the block to
cool before switching the power off.

For information on automating your testing, contact Hart Scientific, Inc.

10.1

Comparison Methods

10.1.1

Direct Calibration

Direct calibration involves testing a probe directly against the dry-well’s tem-
perature display. The method provides the advantage of being quick and easy.

Insert the probe to be calibrated into the well of the dry-well calibrator. The
probe should fit snugly into the calibrator probe sleeve yet should not be so
tight that it cannot be easily removed. Avoid any dirt or grit that may cause the
probe to jam into the sleeve. Best results are obtained with the probe inserted to
the full depth of the well. Once the probe is inserted into the well, allow ade-
quate stabilization time to allow the test probe temperature to settle as de-
scribed. Once the probe has settled to the temperature of the well, it may be
compared to the calibrator display temperature. The display temperature should
be stable to within 0.1°C for best results.

Caution: Never introduce any foreign material into the probe hole of the

insert. Fluids, etc., can leak into the calibrator causing damage to the cal-
ibrator or binding and damage to your probe.

10.1.2

Comparison Calibration

Comparison calibration involves testing a probe against a similar reference
probe. The advantage to this method is that better accuracy can be achieved
since errors due to dry-well inaccuracy, stem effect, and drift can be reduced.

After inserting the probes to be calibrated, allow sufficient time for the probes
to settle and the temperature of the dry-well to stabilize.

Both the reference probe and the probe under test should be the same size and
construction. Using probes with different lengths, diameters and materials will
have different stem effects causing an unknown temperature difference. All
dry-wells have horizontal and vertical gradients that change with temperature .
This is an unknown variable which can be factored out if probes are the same

9105/9107

51

10 Test Probe Calibration

Advertising
This manual is related to the following products: