9 digital communication interface 43, 1 serial communications 43, 1 wiring 43 – Fluke 9107 User Manual

Page 48: 2 setup 43, Figure 9 serial cable wiring diagram 43, 9 digital communication interface, Serial communications, 1 wiring 9.1.2 setup, Figure 9, Serial cable wiring diagram

Advertising
background image

9

Digital Communication Interface

The dry-well calibrator is capable of communicating with and being controlled
by other equipment through the digital interface. Two types of digital interface
are available — the RS-232 serial interface and the IEEE-488 GPIB interface.

With a digital interface the instrument may be connected to a computer or other
equipment. This allows the user to set the set-point temperature, monitor the
temperature, and access any of the other controller functions, all using remote
communications equipment.

9.1

Serial Communications

The calibrator is installed with an RS-232 serial interface that allows serial
digital communications over fairly long distances. With the serial interface the
user may access any of the functions, parameters and settings discussed in Sec-
tion 8 with the exception of the BAUD rate setting.

9.1.1

Wiring

The serial communications ca-
ble attaches to the calibrator
through the DB-9 connector at
the back of the instrument. Fig-
ure 9 shows the pin-out of this
connector and suggested cable
wiring. To eliminate noise, the
serial cable should be shielded
with low resistance between the
connector (DB9) and the shield.

9.1.2

Setup

Before operation the serial in-
terface must first be set up by
programming the BAUD rate
and other configuration parame-
ters. These parameters are pro-
grammed within the serial
interface menu. The serial in-
terface parameters menu is out-
lined in Figure 6.

To enter the serial parameter
programming mode first press

9105/9107

43

9 Digital Communication Interface

RS-232 Cable Wiring for

IBM PC and Compatibles

1 NC
2 RxD
3 TxD
4 NC
5 GND
6 NC
7 RTS
8 CTS
9 NC

2 TxD
3 RxD
4 RTS
5 CTS
6 DSR
7 GND
8 DCD
20 DTR

Instrument

Connector

(DB 9-Pin)

Computer (DTE)

Connector

(DB 25-Pin)

1 NC
2 RxD
3 TxD
4 NC
5 GND
6 NC
7 RTS
8 CTS
9 NC

1 DCD
2 RxD
3 TxD
4 DTR
5 GND
6 DSR
7 RTS
8 CTS
9 NC

Instrument

Connector

(DB 9-Pin)

Computer (DTE)

Connector

(DB 9-Pin)

Figure 9 Serial Cable Wiring Diagram