reaches the new set-point. With the scan disabled the dry-well will heat or cool
at the maximum possible rate.

8.4.1

Scan Control

The scan is controlled with the scan on/off function that appears in the main
menu after the set-point function.

ScAn=OFF

Scan function off

Press “UP” or “DOWN” to toggle the scan on or off.

ScAn=On

Scan function on

Press “SET” to accept the present setting and continue.

S

Accept scan setting

8.4.2

Scan Rate

The next function in the main menu is the scan rate. The scan rate can be set
from .1 to 100 °C/min. The maximum scan rate however is actually limited by
the natural heating or cooling rate of the instrument. This is often less than 100
°C/min, especially when cooling.

The scan rate function appears in the main menu after the scan control function.
The scan rate units are in degrees per minute, degrees C or F depending on the
selected units.

Sr= 10.0

Scan rate in °C/min

Press “UP” or “DOWN” to change the scan rate.

Sr= 2.0

New scan rate

Press “SET” to accept the new scan rate and continue.

S

Accept scan rate

8.5

Temperature Display Hold

The instrument has a display hold function which allows action of an external
switch to freeze the displayed temperature and stop the set-point from scan-
ning. This is useful for testing thermal switches and cut-outs. This section ex-
plains the functions available for operating the temperature hold feature. An
example follows showing how to set up and use the hold feature to test a
switch.

9105/9107

27

8 Controller Operation