4 setting the temperature 14, Setting the temperature, 4 setting the temperature – Fluke 9107 User Manual

Page 20

Advertising
background image

“POWER” switch. The dry-well will turn on and begin to heat to the previously
programmed temperature set-point. The front panel LED display will indicate
the actual dry-well temperature.

The 9107 is field switchable between 115 V and 230 V. Refer to Section 7.2,
Switching to 230 V Operation, for information on switching the voltage.

5.4

Setting the Temperature

Section 8.3 explains in detail how to set the temperature set-point on the cali-
brator using the front panel keys. The procedure is summarized here.

1.

Press “SET” twice to access the set-point value.

2.

Press “UP” or “DOWN” to change the set-point value.

3.

Press “SET” to program in the new set-point.

4.

Press “EXIT” to return to the temperature display.

When the set-point temperature is changed the controller will switch the well
heater on or off to raise or lower the temperature. The displayed well tempera-
ture will gradually change until it reaches the set-point temperature. The well
may require 5 to 10 minutes to reach the set-point depending on the span. An-
other 5 to 10 minutes is required to stabilize with ±0.1°C of the set-point. Ulti-
mate stability may take 15 to 20 minutes more of stabilization time.

14

Hart Scientific

5 Quick Start

Advertising
This manual is related to the following products: