9 Digital Communication Interface

Command Description

Command
Format

Command
Example

Returned

Returned
Example

Acceptable
Values

Display Temperature

Read current set-point

s[etpoint]

s

set: 9999.99 {C or F}

set: 150.00 C

Set current set-point to n

s[etpoint]=n

s=450

Instrument Range

Read scan function

sc[an]

sc

scan: {ON or OFF}

scan: ON

Set scan function:

sc[an]=on/of[f]

ON or OFF

Turn scan function on

sc[an]=on

sc=on

Turn scan function off

sc[an]=of[f]

sc-of

Read scan rate

sr[ate]

sr

srat: 999.99 {C or F}/min

srat: 10.0 C/min

Set scan rate to n degrees per minute

sr[ate]=n

sr=5

.1 to 100°C

Read display temperature hold status

hm[ode]

hm

hm: {OFF or AUTO or NO or NC}

Set display temperature hold mode:

hm[ode]=OF[F]/AU[TO]/N
O/NC

OFF or AUTO or NO
or NC

Set hold mode to off

hm[ode]=OF[F]

hm=of

Set hold mode to automatic

hm[ode]=AU[TO]

hm=auto

Set hold mode to normally open

hm[ode]=NO

hm=no

Set hold mode to normally closed

hm[ode]=NC

hm=nc

Read temperature

t[emperature]

t

t: 9999.99 {C or F}

t: 55.69 C

Read hold status

ho[ld]

ho

ho:{Closed or Open}, 9999.99 {C
or F}

ho: Open, 75.0 C

Secondary Menu

Read proportional band setting

pr[op-band]

pr

pb: 999.9

pb: 15.9

Set proportional band to n

pr[op-band]=n

pr=8.83

Depends on
Configuration

Read cutout setting

c[utout]

c

c: 9999 {C or F}

c: 620 C, in

Set cutout setting:

c[utout]=n/r[eset]

Set cutout to n degrees

c[utout]=n

c=500

Temperature Range

Reset cutout now

c[utout]=r[eset]

c=r

Read heater power

(duty cycle)

po[wer]

po

p%: 9999

po: 1

Ramp and Soak Menu

Read number of programmable set-points

pn

pn

pn: 9

pn: 2

Set number of programmable set-points to n pn=n

pn=4

1 to 8

Read programmable set-point number n

psn

ps3

psn: 9999.99 {C or F}

ps1: 50.00 C

Set programmable set-point number n to n

psn=n

ps3=50

1 to 8, Instrument
Range

Read program set-point soak time

pt

pt

ti: 999

ti: 5

Set program set-point soak time to n minutes pt=n

pt=5

0 to 500

Read program control mode

pc

pc

prog: {OFF or ON}

prog: OFF

Set program control mode:

pc=g[o]/s[top]/c[ont]

GO or STOP or CONT

Start program

pc=g[o]

pc=g

Stop program

pc=s[top]

pc=s

Continue program

pc=c[ont]

pc=c

Read program function

pf

pf

pf: 9

pf: 3

Table 2 Communications Commands