2 set-point value 26, 3 temperature scale units 26, 4 scan 26 – Fluke 9107 User Manual

Page 31: 2 set-point value 8.3.3 temperature scale units, Scan, 4 scan

Advertising
background image

To change the set-point memory press “UP” or “DOWN”.

3. -10.0

New set-point memory 3, –10.0°C

Press “SET” to accept the new selection and access the set-point value.

S

Accept selected set-point memory

8.3.2

Set-point Value

The set-point value may be adjusted after selecting the set-point memory and
pressing “SET”. The set-point value is displayed with the units, C or F, at the
left.

C -10.00

Set-point 3 value in °C

If the set-point value need not be changed then press “EXIT” to resume dis-
playing the well temperature. Press “UP” or “DOWN” to adjust the set-point
value.

C -12.00

New set-point value

When the desired set-point value is reached press “SET” to accept the new
value and access the temperature scale units selection. If “EXIT” is pressed in-
stead then any changes made to the set-point will be ignored.

S

Accept new set-point value

8.3.3

Temperature Scale Units

The temperature scale units of the controller maybe set by the user to degrees
Celsius (°C) or Fahrenheit (°F). The units are used in displaying the well tem-
perature, set-point, and proportional band.

Press “SET” after adjusting the set-point value to change display units.

Un= C

Scale units currently selected

Press “UP” or “DOWN” to change the units.

Un= F

New units selected

8.4

Scan

The scan rate can be set and enabled so that when the set-point is changed the
dry-well will heat or cool at a specified rate (degrees per minute) until it

26