Table 2 communications commands continued 48, Table 2, Communications commands continued – Fluke 9107 User Manual

Page 53: Configuration menu

Advertising
background image

48

Hart Scientific

9 Digital Communication Interface

Command Description

Command
Format

Command
Example

Returned

Returned
Example

Acceptable
Values

Set program function to n

pf=n

pf=2

1 to 4

Configuration Menu

Probe Menu

Read R0 calibration parameter

r[0]

r

r0: 999.999

r0: 100.578

Set R0 calibration parameter to n

r[0]=n

r=100.324

98.0 to 104.9

Read ALPHA calibration parameter

al[pha]

al

al: 9.9999999

al: 0.0038573

Set ALPHA calibration parameter to n

al[pha]=n

al=0.0038433

.00370 to .00399

Read DELTA calibration parameter

de[lta]

de

de: 9.99999

de: 1.46126

Set DELTA calibration parameter to n

de[lta]=n

de=1.45

0.0 to 2.9

Read BETA calibration parameter

be[ta]

be

be: 9.999

be: 0.342

Set BETA calibration parameter to n

be[ta]=n

be=0.342

–100.0 to 100.0

Operating Parameters Menu

Set temperature units:

u[nits]=c/f

C or F

Set temperature units to Celsius

u[nits]=c

u=c

Set temperature units to Fahrenheit

u[nits]=f

u=f

Read cutout mode

cm[ode]

cm

cm: {xxxx}

cm: AUTO

Set cutout mode:

cm[ode]=r[eset]/a[uto]

RESET or AUTO

Set cutout to be reset manually

cm[ode]=r[eset]

cm=r

Set cutout to be reset automatically

cm[ode]=a[uto]

cm=a

Read approach setting

ap[proach]

ap

ap:9

ap:5

Set approach setting to n degrees

ap[proach]=n

ap=15

0 to 20°C

Read stability

ts

ts

ts:9.9

ts:0.5

Set soak stability to n degrees

ts=n

ts=.1

.01 to 4.99°C

Serial Interface Menu

Read serial sample setting

sa[mple]

sa

sa: 9

sa: 1

Set serial sampling setting to n seconds

sa[mple]=n

sa=0

0 to 4000

Set serial duplex mode:

du[plex]=f[ull]/h[alf]

FULL or HALF

Set serial duplex mode to full

du[plex]=f[ull]

du=f

Set serial duplex mode to half

du[plex]=h[alf]

du=h

Set serial linefeed mode:

lf[eed]=on/of[f]

ON or OFF

Set serial linefeed mode to on

lf[eed]=on

lf=on

Set serial linefeed mode to off

lf[eed]=of[f]

lf=of

Calibration Menu (WARNING – changing the following calibration values may change the accuracy of the instrument.)

Read B0 calibration parameter

*b0

*b0

b0: 9

b0: 0

Set B0 calibration parameter to n

*b0=n

*b0=0

–999.9 to 999.9

Read BG calibration parameter

*bg

*bg

bg: 999.99

bg: 15625

Set BG calibration parameter to n

*bg=n

*bg=156.25

–999.9 to 999.9

Communications Commands Continued