Problem

Causes and Solutions

Temperature readout is not the
actual temperature of the well

With the unit stable, slowly rotate the unit. If no change occurs, the unit
may need to be calibrated. Contact Hart Scientific Customer Service. If the
display changes more than twice the normal display deviation, another unit
in the area could be emitting RF energy. Move the unit to a different loca-
tion and rotate the unit again. If the temperature is correct in this new area
or deviates differently than the first area, RF energy is present in the room.
If you have to perform the test in the effected area, use the comparison test
to eliminate any possible errors.

13.2

Comments

13.2.1

EMC Directive

Hart Scientific’s equipment has been tested to meet the European Electromag-
netic Compatibility Directive (EMC Directive, 89/336/EEC). Selection of Light
Industrial of Heavy Industrial compliance has been based on the intended use
of the instrument. Units designed for use in a calibration laboratory have been
tested to Light Industrial Standards. Units designed to be used in the “field”
have been tested to both Light Industrial and Heavy Industrial Standards. The
Declaration of Conformity for your instrument lists the specific standards to
which the unit was tested.

13.2.2

Low Voltage Directive (Safety)

In order to comply with the European Low Voltage Directive (73/23/EEC),
Hart Scientific equipment has been designed to meet the IEC 1010-1 (EN
61010-1) and the IEC 1010-2-010 (EN 61010-2-010) standards.

60

Hart Scientific

13 Troubleshooting

Advertising
This manual is related to the following products: