11 OPTIONS

Part No 1100869

59

Invacare®9000 Topaz™Wheelchair

Repositioning Upper Mounting Bracket

1.

With the telescoping I.V. rod in the lowest position, align the hole of the outer tube with the phillips screw of the inner tube.

2.

Remove phillips screw and set aside.

3.

Remove the inner tube from the outer tube.

4.

Firmly grasp the height adjustment assembly and pull to remove from the outer tube.

5.

Slide the upper mounting bracket up off the outer tube.

6.

Rotate the upper mounting bracket 180° and reinstall onto the outer tube.

7.

Reverse STEPS 1-5 to reassemble the telescoping I.V. rod.

For this procedure, refer to FIGURE 7 on page 60.

This adjustment is required if the telescoping I.V. rod must be mounted on the right side of the wheelchair.