Chapter 3 menus, 2 next menu screens – SIGMA SQC-122c User Manual

Page 37

Advertising
background image

Chapter 3

Menus

3-4

3.2 Next Menu Screens

The Next Menu SoftKey provides access to two sub-menus of the Main Menu that
control display and operating mode preferences.

Next Menu Screens

Zero

Zeros the thickness reading. Useful for resetting or extending
the current deposition layer.

Auto /
Manual

Toggles between Auto and Manual power control. When
Auto/Manual is shown, output power is set by the SQC-122c to
achieve the programmed deposition rate. When Manual/Auto
is shown, the control knob sets the output power.

Next
Graph

Cycles through options for the main display. Choose between
Rate, Rate Deviation, and Power graphs.

Switch
Displays

Toggles between data display options at the bottom of the
SQC-122c screen. Note the differences between the screens
above, and the Main Menu screen in the previous section.

Sensor Info

Replaces the graph with the sensor information screen:

THICKNESS(kA)

0.000

RATE(A/s)

0.0

POWER(%)

0.0

PHASE: Stopped

100.0

50.0

0.0

0.0

25.0

50.0

Power ( % vs. Time)

Next

Menu

Next

Graph

Switch

Displays

Start

12.5

37.5

Sensor

Info

Run #: 1

5-16-00

8:22:39

Source: 1

Max Pow:: 100.00 %

Slew: 100.00 %

THICKNESS(kA)

0.000

RATE(A/s)

0.0

POWER(%)

0.0

PHASE: Stopped

100.0

50.0

0.0

0.0

25.0

50.0

Power (% vs. Time)

Next

Menu

Zero

Auto /

Manual

Start

12.5

37.5

Run #: 1

5-16-00

8:22:39

Source: 1

Max Pow:: 100.00 %

Slew: 100.00 %

THICKNESS(kA)

0.000

RATE(A/s)

0.0

POWER( %)

0.0

Next

Menu

Sensor #

Enable

Frequency

Life

1

2

ON

ON

5.543210 MHz

5.654321 MHz

54.32 %

65.43 %

Advertising