Lake Shore Model 647 Magnet Power Supply User’s Manual

Setup & Connections

2-5

2.5.6 MPS Analog Current And Voltage Monitoring Connections

The MPS provides amplified and buffered current and voltage monitor output signals at the terminal block on

the back panel. Connect these signals to external meters to indicate output current and voltage. Obtain the

Current Monitor signal through connections to terminals 9 (Im) and 11 (m) with positive output currents

producing a positive monitor voltage of 10 mV/A from Im to m.

Obtain the Voltage Monitor signal through connections to terminals 10 (Vm) and 11 (m) with positive terminal

voltages producing a positive monitor voltage of 10 mV/V from Vm to m.

Table 2-4

Analog Monitoring, Programming, & Remote Sense Connections

TERMINAL LABEL

DEFINITION

9

Im

Output Current Monitor – Voltage output from Im to

GND(M) is ±10 mV/A.

10

Vm

Output voltage monitor – Voltage output from Vm to

GND(M) is ±10 mV/V.

11

m

Monitor and program ground. GND(m).

12

Vp

Not Used.

13

+Vs

Not Used.

14

–Is

Negative voltage supply for programming external

current with a potentiometer

15

Ip

Current programming input voltage. Voltage input from

Ip to GND(m) produces ±100 A/V. Voltage may come

from a voltage source or from the center tap of a

potentiometer connected from -Is to +Is.

16

+Is

Positive voltage supply for programming external

current with a potentiometer.

17

18

–S

+S

Remote voltage sense correction.
Correction for load lead drops of up to 0.5 V per lead.

Figure 2-3

Analog Monitoring , Programming, & Remote Sense Connections

2.5.7 External Current Programming

Remotely program MPS output current by an external voltage or potentiometer. Enable external analog

programming via the rear panel I MODE switch. When the I MODE switch is in the INT I position, external

current mode is disabled. When the I MODE switch is in the EXT I position, the external programming voltage

is summed with the internal programming voltage. Set the internal programming to zero for external

programming only. Apply an external voltage from lp to m of 0 to 1.25 volts or use a 10 K

potentiometer to

control the output current over the entire range. Make connections to rear panel detachable terminal block

defined in Tables 2-3 and 2-4 and Figures 2-2 and 2-3. The MPS produces 100 A of output current for 1 V at

the current programming input.

NOTE: MPS protection circuits reduce the effect of open external programming leads. An open external

programming lead forces external programming voltage to approximately 0 volts.