Lake Shore Model 647 Magnet Power Supply User’s Manual

Operation

3-4

3.7 FUNCTION MENU 3 SCREEN
Press Function Menu to display Function Menu 1.
Press Next Menu twice to display Function Menu 3.

EXIT MENU

Returns to the display screen the Function Menu 3

was entered from.

CURRENT ZERO

Enters the Current Zero screen. See Paragraph 3.7.1.

I STEP LIMIT

Enters the Current Step Limit screen. See Paragraph 3.7.2.

3.7.1 Current Zero Screen
Remotely program MPS current and voltage by external voltages or potentiometers in addition to internal

digital programming. Enable external analog programming via the rear panel MODE switches. When the

mode switch is in the INT I or V position, external current or voltage programming mode is disabled. When it

is in the EXT I or V position, external programming voltage sums with internal programming voltage.

There may be configurations, like multiple auto-parallel MPS units, that introduce an output current offset from

zero. This small offset current may translate into a large energy stored in the magnet load. The output current

zero feature allows users to zero this output current offset from the MPS front panel. The current zero feature

is enabled when either internal or external programming is used. The current zero value stores in non-volatile

MPS memory.

Access output current zero from the MPS front panel.

Enter a current setting of zero and allow the output

current to settle to the offset value. Press the

CURRENT ZERO Function key on Function Menu 3 to

display the Current Zero screen with the output

values. The Up or down Numeric Entry keys toggle

the status. If the status is ON, it must be toggled to

OFF to clear the old value before a new value is accepted. When the status is toggled ON, the output current

stores as the current zero and displays as the I ZERO value.

3.7.2 Current Step Limit Screen
When a superconducting magnet quench occurs, the magnet becomes resistive. The output current forces

MPS output to the voltage setting (constant voltage mode) and output current drops rapidly. To avoid

excessive cryogen boil off, the output current setting should be changed to 0 as quickly as possible.

The output current step limit feature allows users to
enter a current step limit which, if exceeded,
automatically resets the output settings to 0A and 1V.
Press the I STEP LIMIT Function Key on Function
Menu 3 to display the Current Step Limit screen.

Use the Cursor keys to move the line indicator up and

down. The ¥ icon indicates parameters that can only be changed using the up or down Numeric Entry keys.

Change all others using all the numeric entry modes, including the cursor. (See Section 3 for keypad

operations.) Below are valid Output Current Step Limit entries:

STATUS:

<OFF> or <ON>

Output Current Step Limit Status. Use the Up or down Numeric Entry keys to toggle the status.

I STEP + 0.00 A to + 999.99 A

Output Current Step Limit. Use any numeric entry mode to change the value. If the MPS detects a change in

output current that exceeds the step limit, it enters Step Limit Mode and forces output settings to 0 amps and

1 volt. On the next update cycle, it closes the FLT contacts to indicate the fault and the internal audio indicator

beeps about once per second. See Chapter 5 for error messages and action to take if the Current Step Limit

is exceeded.

3/ 3

EX IT M E NU

CU R R E N T ZE R O

I ST EP L IM IT

¤

IM A X S E T :+ 7 2 .0 0 A

V M A X S E T :+ 3 2 .0 0 V

S E T T IN G S :

+ 0 .0 0 A

1.0 0 V

: <O F F >

: + 0 .0 0

: IN T E R N A L

- 0 .0 0 A

+ 0 .0 0 V

L O C

C U R R E N T Z E R O :

¤ S T A T U S

I Z E R O

I M O D E

O U T P U T :

¥

*

: <O F F >

: + 10 .0 0 A

- 0 .0 0 A

+ 0 .0 0 V

L O C

C U R R E N T S T E P L IM IT :

¤ S T A T U S

I S S T E P

O U T P U T :

¥

*