Lake Shore Model 647 Magnet Power Supply User’s Manual

Operation

3-2

3.3 GENERAL DISPLAY DESCRIPTION
In general, screen displays are split into two sections. The left section is the Entry Window where users enter

new parameters. The right section is the Menu Window where Menus display when Output values do not.

The

arrow (¤) symbol is the line indicator. The cursor up and down keys move the line indicator.

Shown to the right is the Normal Display screen. It is
unique because it allows Settings entry at any time.
The I

MAX

SET and V

MAX

SET values are the soft

current and voltage setting limits. Settings entered
cannot exceed these limits. These limits can be
changed from the SETUP screen.

The OUTPUT screen indicates the output values and the interface status. The Normal Display key returns the

display to this screen at any time.

3.4 NUMERIC

ENTRY

Move the line indicator to the desired line to change a value. Use the Up or down Numeric Entry keys (not
cursor keys) to increment or decrement the value. Enter accepts the update, while Esc returns to the
previous value. Use the cursor right and left keys to move the cursor to a particular digit, or enter numbers
with the up or down Numeric Entry keys.

3.5 FUNCTION MENU 1 SCREEN

Press Function Menu to display Function Menu 1.
The 1/3 in the upper right corner indicates Menu 1 of
3. To exit Menu 1, press Function Menu again.

3.5.1 Setup

Screen

To verify setup, press the SETUP Function key. The
Setup screen to the right displays with Output values.

Use the Cursor keys to move the line indicator up or

down. The up and down arrow icon indicates

parameters that can only be changed with the up or

down Numeric Entry keys. All others can be changed

using all the numeric entry modes, including the cursor. Below are valid Setup entries:

IEEE/SIO:

See Paragraph 4.1.1 to set the optional IEEE/SIO Interface. <NP> indicates the IEEE/SIO option is Not

Present.

CB DEF: 9600,o,7,1

The Control Bus Definition is fixed at 9600 Baud, odd parity, 7 data bits, and 1 stop bit. See Appendix B.

CB ADD: 0 to 32

Control Bus Address in the range of 0 to 32. See Appendix B.

VW ANGLE: 2 to 9

Display viewing angle. 9 = view from above, 5 = view in the middle, 2 = view from below. Initial condition is 5.

Use the up or down Numeric Entry keys to increment or decrement the value.

MF ID: 647 (1)

MPS ID. 1 = single MPS, 2 = multiple MPS configuration. The MPS and quantity are determined from polling

done at power up. This line is skipped in the cursor up and down line selection.

I

MAX

SET: + 0.00 A to +72.00 A

Soft current limit. A current setting cannot exceed this limit. Initial condition is +72.00 A. Use any numeric

entry mode to change the value. See Chapter 5 for error messages and action to take if IMAX is exceeded.

V

MAX

SET: + 0.00 V to +32.00 V

Soft voltage limit. A voltage setting cannot exceed this limit. Initial condition is +32.00 V. Use any numeric

entry mode to change the value. See Chapter 5 for error messages and action to take if VMAX is exceeded.

- 0 .0 0 A

+ 0 .0 0 V

L O C

¤

IM A X S E T :+ 7 2 .0 0 A

V M A X S E T :+ 3 2 .0 0 V

O U T P U T :

*

S E T T IN G S :

+ 0 .0 0 A

1.0 0 V

1/3

SE T U P

OU T P U T ON L Y

DIS P LA Y P LO T

RA M P S T AT U S

¤

IM A X S E T :+ 7 2 .0 0 A

V M A X S E T :+ 3 2 .0 0 V

S E T T IN G S :

+ 0 .0 0 A

1.0 0 V

S E T U P :

¤ IE E E / S O I

C B D E F

C B A D D

V W A N G L E

M F ID

IM A X S E T

V M A X S E T

O U T P U T :

:

: 9 6 0 0 ,o ,7 ,1

: 0

: 5

: 6 4 7 (1)

: + 7 2 .0 0 A

: + 3 2 .0 0 V

¥

¥

¥

*

- 0 .0 0 A

+ 0 .0 0 V

L O C