Lake Shore Model 647 Magnet Power Supply User’s Manual

Setup & Connections

2-8

8. Verify all connections as summarized in Table 2-5.

9. To add a third MPS or a fourth MPS in parallel, make the RI/FLT,

OVP, CMP and output connections in parallel with the connections of

MPS 1 and MPS 2. For MPS 3, set the CAL AND ID switches on the

rear panel as follows : switches 5 through 8 are OFF, switches 4, 3, 2

and 1 are ON, OFF, ON, ON respectively. For MPS 4, set the rear

panel CAL AND ID switches as follows : switches 5 through 8 are

OFF, switches 4, 3, 2 and 1 are ON, ON, OFF, OFF respectively.

Configure the CMP switches for MPS 3 and MPS 4 as those in MPS 1

and MPS 2.

NOTE: For proper operation, turn on MPS 1 and then the remaining

MPS units within 15 seconds.

1

2

3

4

5

6

7

8

1

2

3

4

5

6

7

8

Figure 2-5

Typical Multiple MPS Connections For Two MPS Units

Table 2-5

Two-MPS Autoparallel

Configuration Connections

MPS 1

MPS 2

Control Bus
8 (OVP)
1 (+RI), 3 (+FLT)
2 (–RI), 4 (–FLT)
+OUT
–OUT

Control Bus
8 (OVP)
1 (+RI), 3 (+FLT)
2 (–RI), 4 (–FLT)
+OUT
–OUT