Appendix a: troubleshooting 89, Problem t emperature monitoring, Cause solution – Phonetics 1108 User Manual

Page 101

Advertising
background image

Appendix A: Troubleshooting

89

The temperature sensor has been

disconnected or has broken wires.

2.

The temperature reading is -20

° F

or -30

° C.

Problem

T

emperature Monitoring:

Examine the wires to temperature sensor and

connect or replace wiring.

T

ry mo

ving the unit to a dif

ferent location.

a)

T

emperature sensing may be af

fected by

a source of ambient heat (ie., direct

sunlight, or heat duct proximity).

4.

T

emperature reading is inaccurate.

3.

T

emperature reads 150

° F

or 65

° C.

T

emperature sensor wires are touching or

have shorted.

V

erify and correct wiring.

b)

T

emperature may require calibration.

After moving or placing the unit away from

ambient heat sources, the temperature may

be calibrated to of

fset inaccurate normal

reading by several degrees. Refer to

Chapter 5, Section 5.6.

Cause

Solution

1.

Can

’t program temperature limits;

or the unit w

on

’t read the

temperature sensor

.

The input isn

’t conf

igured to read a

temperature sensor

.

Press SET and CONFIGURE to program the

input. (See Section 5.1.1 for more informa-

tion on configuring inputs.)

c)

The unit is using the wrong temperature

scale (F

ahrenheit vs. Celsius).

V

erify temperature scale. See Chapter 5,

Section 5.5.