Technical Data 

MI 

5 

Temperature Resolution 

LT 

± 0.1 K (± 0.2°F)* 

 

± 0.25 K (± 0.5°F)** 

* For a zoomed temperature span of 300°C (600°F) 
** For the full temperature range of the unit 

Temperature Coefficient 

MIC 

± 0.05 K per K or ± 0,05% / K whichever is 
greater, at ambient: 23 to 125°C (73 to 185°F) 

MIH 

± 0.05 K per K or ± 0,05% / K whichever is 
greater, at ambient: 23 to 180°C (73 to 356°F) 

MIC, MIH 

± 0.1 K per K or ± 0.1% per K whichever is 
greater, at ambient: 0 to 23°C (32 to 73°F) 

MID 

± 0.15 K per K or ± 0.15% per K whichever is 
greater, at ambient: 0 to 85°C (32 to 185°F) 

Box 

± 0.1 K per K or ± 0.1% per K whichever is 
greater 

Thermal Shock (within 20 min.) 

LT 

± 3.5 K at ΔT 

ambient

 = 25 K (45°F) 

at target temperature of 50°C (45°F) 

Emissivity 

All models 

0.100 to 1.100 

Transmission 

All models 

0.100 to 1.000 