2 emissivity of target object, 3 ambient temperature, 4 atmospheric quality – RayTek MI Miniature Infrared Sensor User Manual

Page 20: Basics, 12 mi 4.2 emissivity of target object

Advertising
background image

Basics

 

12 

MI 

4.2 Emissivity of Target Object 

To  determine  the  emissivity  of  the  target  object  refer  to  section  12.1 

Determination  of  Emissivity

  on  page  81.  If  emissivity  is  low, 

measured  results  could  be  falsified  by  interfering  infrared  radiation 
from  background  objects  (such  as  heating  systems,  flames,  fireclay 
bricks,  etc.  close  beside  or  behind  the  target  object).  This  type  of 
problem can occur when measuring reflective surfaces and very thin 
materials such as plastic films and glass. 
This  measurement  error  can  be  reduced  to  a  minimum  if  particular 
care  is  taken  during  installation,  and  the  sensing  head  is  shielded 
from these reflecting radiation sources. 

4.3 Ambient Temperature 

The  sensing  head  was  developed  for  the  following  ambient 
temperature ranges: 

• MIH: 0 to 180°C (32 to 356°F) 
• MIC:  0 to 125°C (32 to 257°F) 
• MID: 0 to 85°C (32 to 185°F) 

The  MID  can  operate  in  ambient  temperatures  up  to  200°C  (392°F) 
with the air‐cooling accessory. 

4.4 Atmospheric Quality 

If  the  lens  gets  dirty,  infrared  energy  will  be  blocked  and  the 
instrument will not measure accurately. It is good practice to always 
keep  the  lens  clean.  The  Air  Purge  Jacket  helps  keep  contaminants 
from  building  up  on  the  lens.  If  you  use  air  purging,  make  sure  a 
filtered  air  supply  with  clean  dry  air  at  the  correct  air  pressure  is 
installed before proceeding with the sensor installation. 

Advertising